Troubleshooting¶
"No Plate" or "Select Reporter" when adding peaks¶
Open a plate first, and choose the DNA parameter under REPORTER.
Xploid opens old plates at start-up¶
It reopens the plates of the last session. Hold Shift while it starts to skip that.
A plate opens with an old analysis¶
The analysis is restored from INFO\XploidCurrent.xprj. Press LD to go to the saved
default, or New to start again.
The legend did not come in with the plate¶
Xploid looks for <plate name>.csv in the plate folder, then INFO\Zaden.csv. Check the
file name, or use EZ Import on the Legend tab.
Every well shows ploidy 0¶
- No diploid peak or reference 2C is set on Ranges For Ploidy.
- Minimum (%) is too high for this plate.
- The peaks are outside the ranges - check Range (%).
A single well has the wrong ploidy¶
Look at its histogram in the well table. Exclude a wrongly fitted peak by clicking its flag, or move the well with Shift if its peaks are displaced.
The Hybrid tab says "Set Internal Standard please" or "Set Control Peak"¶
Choose both peaks on the Hybrid tab. If it reports Controls not Found, mark control wells with a right-click in the well table, or relax the P-Value.
EZ Export asks for a file name¶
Excel, or the template, was not available, so Xploid falls back to writing a CSV file. Check Excel Template and Sheet Name on the Options tab.
Resetting Xploid¶
Close Xploid and delete %APPDATA%\Xploid\Xploid.ini. Your plates and their analyses are
kept.